Instruction Manual alpha-pH 1000
12
4.3 Control Concept
The main function and sub-function groups are organised in a matrix format as shown below. These functions
can be accessed via the front keypad for configuration and selection of parameters.
CALIBRATION
MEASURE
CODE
OFFSET
SET C
o
SET-POINT 1
SET-POINT 2
CONTROLLER
MEASUREMENT
RANGE
CONFIGURATION
Set Offset
Switch On/Off ATC
Set value 1
Set value 2
Controller type
Current output
0..20 / 4..20 mA
pH / ORP Symm.
/ Asymmetrical
CLn
ALr
Process temp.
(MTC)
Calibrate 1st value
MIN / MAX Function
MIN / MAX Function
N.O/N.C. Contact
pH / ORP
at 0/4 mA
Alarm delay
00
Intv
Calibrat e 2nd value
Calib. temp.
(MTC)
Hysteresis 1
Hysteresis 2
Proportional
range XP
pH / ORP
at 20 mA
Alarm contact
cont./wiper contact
1000
Dur
Display zero point
and slope
Calibrate Pt100
Pick-up delay
contact 1
Pick-up delay
contact 2
Cycle time bzw.
max. frequency
Electrod e t ype
Glass/Antimony
Drop out delay
contact 1
Drop out delay
contact 2
Reset to factory
default settings
The controller offers two levels of password protection: (1) for direct access to calibration function and (2) for
setting or editing specific controller parameters or functions in the SETUP mode to suit individual requirements.
Note: The passwords are not user-defined and have been set by factory. It is very important to keep these
passwords strictly confidential to avoid unauthorised tampering of the system at all times.
Note: If the user reads parameters only, the controller automatically reverts to Measurement mode if no key is
pressed for 30 seconds.