Instruction Manual Salt 6
11
4 CALIBRATION
4.1 Important Information on Meter Calibration
The Salt 6 meter has a special built-in algorithm to linearize for the
measurement of Sodium Chloride concentration. Therefore the specified
accuracies from a One-point calibration anywhere in the measurement range,
specified accuracies can be applied for the entire measurement range.
When you recalibrate your meter, the old calibration will be replaced by the new
calibration point. For example, if you previously calibrated your conductivity
meter at 2.5% and you recalibrate at 1.0, the meter will replace the old
calibration data (2.5%).
To completely recalibrate your meter, or when you use a replacement probe, it
is best to clear all calibration data. To erase all calibration data completely, see
Section 6.5 – Restore Factory Default Values.