11
Press ON/OFF key to power up your meter.
1) The first screen shows [Con 6] (or [tdS 6]) which is the meter’s name.
2) Second screen shows [C 1.0] which is the conductivity cell constant, k.
You can select different cell constant of 0.1, 1.0 or 10.0. Refer to Section
on Advance Setup. Default value is k=1.0.
3) Third screen shows [t 25.0 °C] which is the Normalization Temperature.
You can set Normalization Temperature at either 25 °C or 20 °C. Refer to
Section on Advance Setup. Default value is 25 °C.
4) Fourth screen shows [t 2.1%] which is the Temperature Coefficient. You
can customise the meter with different Temperature Coefficient value
from 0.0 to 3.0 %/°C from the Advance Setup mode. Default value is 2.1
%/°C.
5) All LCD segments will light up for 2 seconds, and change into
measurement mode.
6) You are now ready for conductivity measurement.