3.5 Manual Calibration
In Manual Calibration mode you are not limited to the conductivity calibration standards
listed previously in Table 1. This example shows a manual calibration sequence using a
12.00 mS/cm conductivity calibration standard.
1. If necessary, press MODE key to select conductivity
mode.
1. Rinse the probe thoroughly with de-ionized water or
a rinse solution, then rinse with a small amount of
calibration standard.
1. Dip the probe into the calibration standard. Stir the
probe gently to create a homogeneous sample.
Allow time for the reading to stabilize.
1. Press CAL to enter conductivity calibration mode.
The [CA] indicator will appear briefly, then a value
will appear flashing.
NOTE: To exit calibration without confirmation, press
CAL to return to measurement mode.
2. When the value is stable, press or as needed
to adjust the value to match your calibration
standard.
3. Press ENTER to confirm the adjusted value. [CO]
will appear briefly indicating that the calibration was
successful. The meter returns to measurement
mode.
4. Repeat steps 1-6 as needed with additional
calibration standards.