4. For zooming setting, use the or keys to set
the conductivity low zoom value for the 4 mA
current and press the ENT key to confirm. The
transmitter now displays the high zoom value
setup.
5. Use the or keys again to set the
conductivity high zoom value for the 20 mA
current and press the ENT key to confirm and
return to the range setup mode.
6. Press both and keys together to return to
the measurement mode.